When determining total nitrogen (or protein) of a sample using LECO’s FP828, it is crucial to select the right sample container for the job. Several key considerations must be taken into account, like the sample mass capacities of the container, compatibility of the container with the matrix, ash production, and more.

LECO - News - Choosing the Right Sample - Quick capsules image

Dimensions and Sample Mass Capacities

For those working with food or feed materials, the document “Sample Containers and Crucibles for Organic Analysis” provides valuable insights into the different types of sample containers available, their nominal sample mass capacities, and dimensions. For instance, Quick-Cap Capsules, Tin Capsules, and Tin Foil Cups are commonly used with the FP828 for nitrogen determination. Each container has different dimensions and sample mass capacities, which are crucial for ensuring accurate and reliable results.

Compatibility with the Matrix

Al evaluar la compatibilidad del recipiente de la muestra con la matriz de la muestra, es de suma importancia considerar el contenido de humedad. Las cápsulas Quick-Cap son adecuadas solo para muestras secas y sólidas y no pueden usarse con muestras acuosas o aquellas con alto contenido de humedad. El material de gelatina de la cápsula se disolverá al contacto con materiales acuosos. Por otro lado, las cápsulas de estaño pueden usarse tanto con muestras sólidas como acuosas, lo que las hace más versátiles para diferentes tipos de materiales de alimentos y piensos.

LECO - News - Choosing the Right Sample - Tin capsules image

Ash Production

The choice of container also has an influence on ash production. Quick-Cap Capsules combust completely, resulting in minimal ash production. However, Tin Capsules oxidize to form tin oxide, leading to a higher ash load.

Atmospheric Blank Influence

The atmospheric blank influence is another critical factor to consider. Quick-Cap and Tin Capsules are left open during analysis, allowing any atmosphere present in the capsule to be purged within the purge chamber, reducing the possibility of nitrogen bias. Capsules should not be sealed shut to avoid trapping atmosphere with the sample, which could lead to inaccurate results.

LECO - News - Choosing the Right Sample - Foil cups image

Low Nitrogen Concentrations

In addition to these considerations, there are containers that are highly recommended for materials with low nitrogen concentrations. Large Capsules and Foils are best for these analyses (as well as samples that involve low-density or non-homogenous materials) because the larger capsule size allows for an increased sample mass. Similarly, Large Tin Capsules can accommodate a larger sample mass for both solid and liquid samples. When using a Tin Foil to determine lower levels of nitrogen within a sample, determining an atmospheric blank by analyzing a non-nitrogen containing material and using the average nitrogen value obtained to correct for the trapped atmosphere is recommended.

Conclusion

Overall, selecting the appropriate sample container type for nitrogen (or protein) determination using the FP828 instrument involves careful consideration of the sample type, container dimensions, sample mass capacity, compatibility with the sample matrix, ash production, and atmospheric blank influence. By following the guidelines and recommendations provided in the document “Sample Containers and Crucibles for Organic Analyses”, users can ensure accurate and reliable results for their specific needs. Download this document today for use in your lab!